Hettich HAEMATOKRIT HT1201
At
16,060 g sedimentation is completed in just a little over 6 minutes. At
any rate, a run time of 10 minutes is sufficient. The rotor comes to a
braked standstill. Hematocrit determination with standard capillaries is carried out with 24-place rotor no. 2076. Each capillary is secured within its own chamber and supported by a holding tray. This holding tray serves to contain glass shards and any leakage should a tube break. It is easy to clean and disposable in the case of glass breakage. The lid of rotor 2076 is cover and evaluation disc in one. Capillaries, to be used for quantitative buffy coat analysis are centrifuged in 20-place rotor no. 2056 prior to further analysis. |
Literature | Call 01865 750375 for sales advice |
HAEMATOKRIT HT1201 | |
---|---|
Max. capacity in a hematocrit rotor: |
24 standardcapillaries |
Max. speed (RPM): | 13.000 min-1 |
Max. RCF: | 16.060 |
Dimensions (H x W x D): | 247 x 275 x 330 mm |
Weight: | approx. 10 kg |
Refrigeration: | Air cooling |
Rotors | |
Haematocrit rotor | 24 standard capillaries, max. RCF 16,060 |
Disc rotor | 20
capillaries for buffy coat analysis, max. RCF 16,060 |